MassHunter GC/MS Acquisition Software B.06.01
This revision provides
updates to the Agilent 7200 series Q-TOF GC/MS.
Here is what is new or
changed in this revision:
New B.06.01 Acquisition
Enhancements
In addition to changes from previous versions of
Acquisition, B.06.01 includes
·
7200 series Q-TOF GC/MS
1. TOF Conditioning Panel will start automatically once vacuum has been
reached
2. Bakeout button has been added to the Remove Ion Source Panel
3. Vent and Pump down dialog has been updated to continuously show the status
of the instrument.
4. The autotune log has been changed to not record
spectral data for ramps and quad profiles.
5. Improved CI gas status monitoring and display.
6. Removing Ion Source steps has been improved to handle faults and returning
to a safe state if user aborts.
7. Continuous data acquisition in manual tune.
Operating Systems
·
Chinese Windows
7 (64) Professional SP1, Windows 7 (64) Professional
·
English Windows
7 (64) Professional SP1, Windows 7 (64) Professional
·
Japanese Windows
7 (64) Professional SP1, Windows 7 (64) Professional
Windows Vista Business SP1 is no longer supported. GC/MS Acquisition B.05.02 is the last version that supports Windows Vista
Business SP1.
No other languages or versions of Windows are supported.
MassHunter software requires the Microsoft .NET 4.0 Framework to be
present. See
X:\Supplemental\tools\Framework\ReadMe.txt for more information on installing
it if necessary.
GC and MS Firmware
·
Click here GC and MS Firmware Revisions for information
on current firmware revisions.
·
The GC firmware version must be the appropriate
version listed above or later.
·
The GC firmware is included on the MassHunter GC/MS
Acquisition Software Installation DVD.
·
MassHunter GC/MS Acquisition Software B.06.01 requires
7000 series Triple Quad GC/MS MS firmware revision G.7000.037F.
·
MassHunter GC/MS Acquisition Software B.06.01 requires
7200 series Q-TOF GC/MS MS firmware revision
G.7200.01.09.
·
The MS firmware is included as part of the MassHunter
GC/MS Acquisition Software installation, and is loadable via the Instrument
Configuration program.
·
Information on both current and previous MassHunter
GC/MS Acquisition Software firmware revisions is contained on the Installation
DVD:
X:\CurrentFirmwareRevisions-en.htm
X:\PreviousReadmeMS-en.htm
Instruments Supported
·
7200 series Q-TOF GC/MS
·
7000 Triple Quadrupole GC/MS
·
7890A GC
·
7683 Autosamplers
·
7693 Autosamplers
·
G650x PAL Autosamplers (with G9257AA Agilent CTC Control Software, B.01.03 or higher)
Instruments Not Supported
·
All other GCs and Mass Spec instruments
Software Supported
·
G9257AA Agilent Data System Addon for CTC Control (rev
B.01.03 or higher)
·
G4600AA/G4800AA Agilent Instrument Utility Software
(rev B.01.06 or higher)
Printers and Drivers
·
LaserJet P3005d:
This printer is fully tested and is supported.
The supported driver is PCL
6.
·
Canon LBP3310:
For Japanese Windows, this printer driver is fully tested and is supported
and is included on the Installation DVD.
For the latest drivers, visit the HP website at http://welcome.hp.com/country/us/en/support.html
Information on previous MassHunter GC/MS Acquisition Software revision is contained
on the Installation DVD:
X:\PreviousReadmeMS-en.htm