SSB
(Software Status Bulletin) - Known issues at time of release
SRB
(Software Release Bulletin) - Issues fixed from last release
Release notes - Supplemental information
The following table recommends actions that will help you resolve
common symptoms or conditions that may occur during or after software
installation.
|
Symptom or condition
|
Recommended action
|
|
When
you try to install this software, the installer stops because your computer
does not have the correct version of Agilent I/O Libraries (SICL).
|
Install
Agilent I/O Libraries (SICL) M.01.01
|
|
Added a
6850 or 6890A with JetDirect card J4100 or J2552, with older firmware.
Software will not connect to or communicate with the instrument.
|
Install Agilent BootP Service: Use the BootP server
to assign an IP address to a 6850 or 6890A with JetDirect cards.
|
|
If you
need to update the firmware for your GC and GC/ALS instruments install and
use the GC Firmware Update Utility.
|
Install the GC Firmware Update Utility.
|
Please
visit the Agilent Technologies Chemical/Biochemical web site at www.chem.agilent.com
for:
• Online stores for all Agilent products
• Technical support
for all Agilent GC products
• Literature library

|